Table 2.
Analytical conditions of GC-ECD, GC-NPD and GC-MSD

Detector μECD NPD GC-MSD
Inlet temperature 230°C 210°C 230°C
Detector temperature 280°C 320°C 280°C
Flow rate 1.0 mL/min (N2) 60 mL/min (Air) 1.0 mL/min (He)
1.2 mL/min (N2)
3 mL/min (H2)
Column DB-1701 DB-1701 HP-5MS
Oven temperature 150°C, 1min 150°C, 1min
12°C/min 12°C/min 100°C, 2min
240°C, 2min 200°C, 8min 10°C/min
10°C/min 10°C/min 280°C, 15min
280°C, 12min 260°C, 8min