Table 1. Analytical conditions of HPLC

Parameters Conditions
Instrument HPLC system (Shiseido, Nanospace SI-2)
Detector PDA detector (254 nm)
Column Capcellpak C18 Type MGⅡ, 5 μm, 4.6×250 mm (Shiseido)
Temperature 40°C
Injection volume 10 μL
Flow rate 1.0 mL/min
A: 0.1% formic acid in water
B: Acetonitrile/MeOH (85:15 v/v)
Mobile phase Gradient Time (min) % A % B
0 85 15
24 76 24
30 20 80
32 85 15
40 85 15
Run time 40 min